A RetroSearch Logo

Home - News ( United States | United Kingdom | Italy | Germany ) - Football scores

Search Query:

Showing content from https://pubmed.ncbi.nlm.nih.gov/15662775/ below:

A comparison of the gamma knife model C and the automatic positioning system with Leksell model B

Object: The authors sought to compare the quality of treatment planning, radiation protection, and the time taken for treatment in the Leksell gamma knife model B with that using the model C Automatic Positioning System (APS).

Methods: Data were obtained in 463 patients treated with the B model and 518 patients treated with the C model. Data were analyzed in patients in whom the following diagnoses had been made: vestibular schwannoma, pituitary adenoma, meningioma, solitary metastasis, and other benign and malignant solitary tumors. Patients with arteriovenous malformations, ocular lesions, and functional diagnoses were excluded from this study.

Conclusions: With the C model there was a better conformity for most treated targets, such as vestibular schwannomas (p = 0.005) and meningiomas (p = 0.015). The level of radiation exposures to personnel was significantly decreased when using the model C (p < 0.001). There was no significant difference in radiation exposure of extracranial structures for the same number of shots in patients treated by both models. The mean time saved using the C model with the APS was 41 minutes per treatment. It would seem that the gamma knife model C permits better dose conformity, shorter treatment times, and less radiation exposure to personnel.


RetroSearch is an open source project built by @garambo | Open a GitHub Issue

Search and Browse the WWW like it's 1997 | Search results from DuckDuckGo

HTML: 3.2 | Encoding: UTF-8 | Version: 0.7.3